B-10
SCPI Tables (Abridged)
Model 6220/6221 User’s Manual
Return to
Table B-10
Trigger command summary
Command*
Description
Default
INIT
Trigger initiation:
[:IMM]
Initiate one trigger cycle.
OFF
ABOR
Reset trigger system.
ARM
Arm layer commands:
:SOUR <name>
Select event detector: IMM, TIM, BUS, TLIN, BST, PST, NST, or MAN.
:SIGN
Bypass ARM control source.
:TIM <n>
Set timer interval (seconds): 0 to 99999.99. 1msec resolution.
0.1
:DIR <name>
Control ARM source bypass: SOUR or ACC.
ACC
:ILIN <NRf>
Set ARM input signal line: 1 to 6.
1
:OLIN <NRf>
Set ARM output signal line: 1 to 6.
2
:OUTP <name>
Enable/disable ARM output trigger: TENT, TEX, or NONE.
NONE
TRIG
Trigger layer commands:
:SOUR <name>
Select event detector: IMM or TLIN.
IMM
:SIGN
Bypass TRIG control source.
:DIR <name>
Control TRIG source bypass: SOUR or ACC.
ACC
:ILIN <NRf>
Set TRIG input signal line: 1 to 6.
1
:OLIN <NRf>
Set TRIG output signal line: 1 to 6.
2
:OUTP <name>
Enable/disable TRIG output trigger: SOUR, DEL, or NONE.
NONE
* Arm layer commands control entire sweep. Trigger Layer commands control individual sweep steps.
Table B-11
Units command summary
Command
Description
Default
UNIT
Commands to set reading units for Delta, Pulse Delta, and Differen-
tial Conductance:
[:VOLT]
Volts readings are sent from 2182/2182A to 622x.
[:DC] <name>
Specify reading units*: V, OHMS, W, or SIEM.
V
:POW
Power reading type for Pulse Delta (6221):
[:TYPE] <name>
Set power units reading type: PEAK or AVER.
AVER
* The <name> parameter for Siemens can be sent as S, SIEM, or SIEMENS.
Summary of Contents for 6220 DC
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ......
Page 36: ...1 20 Getting Started Model 6220 6221 User s Manual Return to Section 1 topics...
Page 131: ...6 10 Averaging Filter Math and Buffer Model 6220 6221 User s Manual Return to Section 6 topics...
Page 148: ...A Specifications...
Page 167: ......
Page 169: ......
Page 170: ......