Model 6220/6221 User’s Manual
Output Connections
2-7
Return to
Commands for triax inner shield connection:
Example
– Turns off the current source OUTPUT and connects the inner shield of
the OUTPUT connector to Cable Guard:
OUTPut OFF
OUTPut:ISHield GUARd
Triax output low
Output Low can be connected to Earth Ground (which is the outer shell of the triax
connector) or it can float. By default, Output Low floats.
Perform the following steps to check or change the Output Low connection:
1.
On the Model 622x, press the
TRIAX
key to display the
CONFIGURE
TRIAX
menu.
2.
Using the cursor controls, place the blinking cursor on
OUTPUT LOW
and
press the
ENTER
key to display the
TRIAX OUTPUT LOW
options.
3.
Place the cursor on
FLOATING
or
EARTH GROUND
and press the
ENTER
key.
4.
Press the
EXIT
key to return to the normal display state.
Remote programming (triax output low)
Commands for output low connection:
Example
– Disconnects (floats) triax output low from Earth Ground:
OUTPut:LTEarth OFF
OUTPut:ISHield?
‘ Query connection for triax inner shield.
OUTPut:ISHield <name>
‘
‘
Connect inner shield to Cable Guard or Output Low.
<name> = GUARd or OLOW
OUTPut:LTEarth?
‘ Query connection for output low.
OUTPut:LTEarth <b>
‘
‘
‘
‘
Enable or disable triax output low connection to Earth
Ground.
<b> = ON (Earth Ground)
= OFF (Floating)
Summary of Contents for 6220 DC
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ......
Page 36: ...1 20 Getting Started Model 6220 6221 User s Manual Return to Section 1 topics...
Page 131: ...6 10 Averaging Filter Math and Buffer Model 6220 6221 User s Manual Return to Section 6 topics...
Page 148: ...A Specifications...
Page 167: ......
Page 169: ......
Page 170: ......