Model 6220/6221 User’s Manual
DC Current Source Operation
3-15
Return to
Remote programming – source output commands
lists the commands to configure and control the DC output. A program-
ming example to output DC current is also provided.
1.
To select a fixed source range, specify the current output value that is going to
be sourced. The Model 622x will go to the lowest range that can source that
value. For example, If you are going to source 25mA, let <n> = 25e-3. The
100mA range will be selected.
2.
Selecting a fixed source range disables autorange.
3.
The output must be off in order change the output response. Sending this
command while the output is on will generate error -220 Execution Error.
4.
OUTP OFF turns the output off, but does change the set output level.
SOUR:CLE sets the output level to zero and then turns the output off.
Table 3-2
DC output commands
Command
Description
Default
CLEar
Turns output off and sets output level to zero.
CURRent:RANGe <n>
Sets current source range (amps).
1, 2
<n> = -105e-3 to 105e-3
100e-3
CURRent:RANGe:AUTO <b>
Enables or disables source autorange.
<b> = ON or OFF
OFF
CURRent <n>
Sets DC current source output level (amps).
<n> = -105e-3 to 105e-3
0.0
CURRent:COMPliance <NRf>
Sets voltage compliance (volts).
<NRf> = 0.1 to 105
10.0
CURRent:FILTer <b>
Enables or disables the output analog filter.
<b> = ON or OFF
OFF
OUTPut:RESPonse <name>
Select fast or slow output response speed for 6221.
<name> = FAST or SLOW
3, 4
FAST
OUTPut <b>
Turn output on or off.
<b> = ON or OFF
OFF
SourceMeter
Sets output to zero, then turns the output off
4
Summary of Contents for 6220 DC
Page 2: ......
Page 4: ......
Page 6: ......
Page 16: ......
Page 36: ...1 20 Getting Started Model 6220 6221 User s Manual Return to Section 1 topics...
Page 131: ...6 10 Averaging Filter Math and Buffer Model 6220 6221 User s Manual Return to Section 6 topics...
Page 148: ...A Specifications...
Page 167: ......
Page 169: ......
Page 170: ......