4-8
Sweeps
Model 6220/6221 User’s Manual
Return to
Remote sweep operation
Procedures for programming and running a sweep are shown on the following
pages. Each of these procedures includes commands for a typical sweep exam-
ple.
summarizes parameters for each of these examples. See
for listings of sweep commands.
Table 4-2
Sweep example parameters
Sweep type
Parameters for sweep examples
Linear staircase sweep (
Start current: 1mA
Stop current: 10mA
Step current: 1mA
Delay: 1s
Source range: best fixed
Compliance abort: off
Sweep count: 1
Bias current: 100
μ
A
Compliance: 10V
)
# points: 5
Points: 3mA, 1mA, 4mA, 5mA, 2mA
Delay: 3s, 1s, 4s, 5s, 2s
Compliance: 3V, 1V, 4V, 5V, 2V
Source range: auto
Compliance abort: off
Sweep count: 1
Bias current: 50
μ
A