Model 6220/6221 User’s Manual
Wave Functions (6221 Only)
7-13
Return to
Programming sine waves
1.
Restore defaults with this command
:
*RST
2.
Configure the waveform
.
Examples
– The following commands configure a 1kHz sine wave with an
amplitude of 10mA, 1mA offset, and phase marker off:
3.
Arm and trigger the waveform, turn on output:
4.
To stop generating the waveform and turn the source output off before the
duration elapses, send this command:
SOUR:WAVE:FUNC SIN
‘ Select sine wave.
SOUR:WAVE:FREQ 1e3
‘ Set frequency to 1kHz.
SOUR:WAVE:AMPL 1e-2
‘ Set amplitude to 10mA.
SOUR:WAVE:OFFS 1e-3
‘ Set offset to 1mA.
SOUR:WAVE:PMAR:STAT OFF
‘ Turn off phase marker.
SOUR:WAVE:DUR:TIME 5
‘ 5 second duration.
SOUR:WAVE:RANG BEST
‘ Select best fixed source range.
SOUR:WAVE:ARM
‘ Arm waveform.
SOUR:WAVE:INIT
‘ Turn on output, trigger waveform.
SOUR:WAVE:ABOR
‘ Stop generating waveform.