4-10
Sweeps
Model 6220/6221 User’s Manual
Return to
Running a custom sweep
1.
Configure source functions
.
Examples
– The following commands restore defaults and set the bias
current to 50
μ
A:
2.
Configure the sweep
.
Examples
– The following commands configure a single custom sweep
with five points, different delay and compliance settings for each point, auto
source range, and compliance abort disabled:
3.
Arm and run the sweep
:
4.
When the sweep is done, turn the source output off with this command:
OUTP OFF
*RST
‘ Restore 622x defaults.
SOUR:CURR 5e-5
‘ Set bias current to 50
μ
A
SOUR:SWE:SPAC LIST
‘ Select custom sweep.
SOUR:SWE:RANG AUTO
‘ Select auto source range.
SOUR:LIST:CURR 3e-3,1e-3,4e-3,
5e-3,2e-3
‘
‘
Sweep points: 3mA, 1mA, 4mA,
5mA, 2mA.
SOUR:LIST:DEL 3,1,4,5,2
‘ Delay points: 3s, 1s, 4s, 5s, 2s.
SOUR:LIST:COMP 3,1,4,5,2
‘ Compliance = 3V, 1V, 4V, 5V, 2V.
SOUR:SWE:COUN 1
‘ Set sweep count to 1.
SOUR:SWE:CAB OFF
‘ Disable compliance abort.
SOUR:SWE:ARM
‘ Arm sweep, turn on output.
INIT
‘ Trigger sweep.