Model 6220/6221 Reference Manual
Sweeps
4-7
Return to
shows an example of a custom sweep over the range of 1mA to 10mA
with arbitrary steps. When the sweep is triggered, the current goes from the bias
level to the first point in the sweep. The unit cycles through the sweep points in the
programmed order. The current output remains at the last step when the sweep is
done. Note that the sweep delay period is different for each step.
Figure 4-4
Custom sweep example (1mA to 10mA, arbitrary steps)
Delay
Delay
Delay
Delay
Start
Bias
Stop
Delay
1
10
Current
(mA)
7.5
5
2.5
Sweep delay can be
different for each step.
0