8-8
Triggering
Model 6220/6221 Reference Manual
Return to
Configuring triggering
The following is a typical example of using trigger configuration to fine-tune
sweeps. For further details on the steps for sweep configuration, see
sweep operation,” on page 4-10
. When setting trigger model parameters, keep the
following in mind:
•
The arm layer controls the entire sweep.
•
The trigger layer controls individual sweep steps.
1.
Press CONFIG then TRIG to enter the trigger configuration menu.
2.
Select ARM-LAYER. Press ENTER, then set your parameters as follows:
a.
ARM-IN: Choose the ARM-IN EVENT, set TIMER interval (if using
timer), TLINK line (1-6), and EVENT DETECT BYPASS (ONCE or
NEVER).
b.
ARM-OUT: Set the LINE (1-6), and EVENTS (TRIG-LAYER-EXIT ON
or OFF, TL-ENTER ON, or OFF).
3.
From the main trigger configuration menu, select TRIGGER-LAYER, then
press ENTER and set your parameters as follows:
a.
TRIGGER-IN: Set the TRIGGER-IN SOURCE, TRIGGER LINK line
(1-6), and EVENT DETECT BYPASS (ONCE or NEVER).
b.
TRIGGER-OUT: Set the trigger LINE (1-6) and events
(SOURCE ON/OFF, DELAY ON/OFF).
4.
Configure and run your sweep as outlined in
“Front panel sweep operation,”
. Depending on how you configured the trigger model, you
may have to apply triggers from the appropriate sources to start the sweep
and to cycle through the sweep steps.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Summary of Contents for 6220
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...