17-18 Calibration
Model 6220/6221 Reference Manual
Return to
Step 3: Calibrate compliance
1.
Connect the digital multimeter (DMM) to the Model 6220/6221 OUTPUT
jack, as shown in
. Set the DMM to measure DC voltage with
auto range enabled.
2.
Send these commands in order to calibrate compliance using the
<DMM_Reading> parameter (actual digital multimeter reading) for each
step:
OUTP ON
‘ Turn on output.
SOUR:CURR:RANG 2e-3
‘ Select 2mA range.
SOUR:CURR 2e-3
‘ 2mA for cal.
SOUR:CURR:COMP 20
‘ Set compliance to +20V
CAL:PROT:SENS <DMM_Reading>
‘ Cal20V using reading.
SOUR:CURR:COMP 0.1
‘ Set compliance to +0.1V
CAL:PROT:SENS <DMM_Reading>
‘ Cal0.1V using reading.
SOUR:CURR -2e-3
‘ Output -2mA for cal.
SOUR:CURR:COMP 20
‘ Set compliance to -20V
CAL:PROT:SENS <DMM_Reading>
‘ Calibrate -20V using reading.
SOUR:CURR:COMP 0.1
‘ Set compliance to -0.1V
CAL:PROT:SENS <DMM_Reading>
‘ Calibrate -0.1V using reading.
SOUR:CURR 2e-3
‘ 2mA for cal.
SOUR:CURR:COMP 100
‘ Set compliance to +100V
CAL:PROT:SENS <DMM_Reading>
‘ Cal100V using reading.
SOUR:CURR:COMP 21
‘ Set compliance to +21V
CAL:PROT:SENS <DMM_Reading>
‘ Cal21V using reading.
SOUR:CURR -2e-3
‘ Output -2mA for cal.
SOUR:CURR:COMP 100
‘ Set compliance to -100V
CAL:PROT:SENS <DMM_Reading>
‘ Calibrate -100V using reading.
SOUR:CURR:COMP 21
‘ Set compliance to -21V
CAL:PROT:SENS <DMM_Reading>
‘ Calibrate -21V using reading.
OUTP OFF
‘ Turn off output.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Summary of Contents for 6220
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...