Table 7-1
Wave function characteristics
Wave function
Characteristics
Sine wave
Frequency: 1mHz to 100kHz (
)
Amplitude: 1pA to 105mA peak (
Range: 2nA to 100mA (
Offset: 0 to ±105mA (
Phase marker: 0° to 360° (
)
Duration: 100ns to 999999.999s (
Square wave
Frequency: 1mHz to 100kHz (
)
Amplitude: 1pA to 105mA peak (
Range: 2nA to 100mA (
Offset: 0 to ±105mA (
)
Phase marker: 0° to 360° (
)
Duration: 100ns to 999999.999s (
Ramp wave
Frequency: 1mHz to 100kHz (
)
Amplitude: 1pA to 105mA peak (
Range: 2nA to 100mA (
Offset: 0 to ±105mA (
)
Phase marker: 0° to 360° (
)
Duration: 100ns to 999999.999s (
Arbitrary wave
Frequency: 1mHz to 100kHz (
)
Amplitude: 1pA to 105mA peak (
Range: 2nA to 100mA (
Offset: 0 to ±105mA (
Number of points: 2 to 65,535
Phase marker: 0° to 360° (
)
Duration: 100ns to 999999.999s (