Model 6220/6221 Reference Manual
Wave Functions (6221 Only)
7-21
Return to
Remote wave function operation
Procedures for programming and generating waveforms for each of the four wave-
form types are given on the following pages. Each of these procedures includes
commands for a typical wave function example.
summarizes parameters
for each of these examples. See
“SCPI commands — wave functions,” on
for details on wave function commands.
NOTE
User setups cannot be saved or recalled while
Wave is armed or running. The *SAV and *RCL
commands will generate error +413 Not al-
lowed with mode armed.
Table 7-3
Waveform example parameters
Waveform
Parameters for waveform examples
Sine wave (
)
Frequency: 1kHz
Amplitude: 10mA
Offset: 1mA
Phase marker: off
Duration: 5s
Square wave (
Frequency: 10kHz
Amplitude: 1mA
Offset: 0
Duty cycle: 50%
Phase marker: 180°, line 1
Duration: 10s
Ramp wave (
Frequency: 50kHz
Amplitude: 5mA
Offset: 0
Duty cycle: 25%
Phase marker: 90°, line 1
Duration: 15s
Arbitrary wave (
Arbitrary points: -0.5, -0.25, 0, 0.3, 0.4
Range: best fixed
Frequency: 100kHz
Amplitude: 25mA
Offset: 0
Phase marker: Off
Duration: 20s
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Summary of Contents for 6220
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...