7-24
Wave Functions (6221 Only)
Model 6220/6221 Reference Manual
Return to
Programming ramp waveforms
1.
Restore defaults with this command:
*RST
2.
Configure the waveform.
Examples
– The following commands configure a 50kHz ramp wave with
an amplitude of 5mA, 0mA offset, 25% duty cycle, 90° phase marker, and
15 second duration:
3.
Arm and trigger the waveform, turn on output:
4.
To stop generating the waveform and turn the source output off before the
duration elapses, send this command:
SOUR:WAVE:FUNC RAMP
‘ Select ramp wave.
SOUR:WAVE:FREQ 5e4
‘ Set frequency to 50kHz.
SOUR:WAVE:AMPL 5e-3
‘ Set amplitude to 5mA.
SOUR:WAVE:OFFS 0
‘ Set offset to 0mA.
SOUR:WAVE:DCYC 25
‘ Set duty cycle to 25%.
SOUR:WAVE:PMAR:STAT ON
‘ Turn on phase marker.
SOUR:WAVE:PMAR 90
‘ Set phase marker to 90°.
SOUR:WAVE:PMAR:OLIN 1
Use line 1 for phase marker.
SOUR:WAVE:DUR:TIME 15
15 second duration.
SOUR:WAVE:RANG BEST
‘ Select best fixed source range.
SOUR:WAVE:ARM
‘ Arm waveform.
SOUR:WAVE:INIT
‘ Turn on output, trigger waveform.
SOUR:WAVE:ABOR
‘ Stop generating waveform.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com