7-26
Wave Functions (6221 Only)
Model 6220/6221 Reference Manual
Return to
Programming an externally triggered waveform
1.
Restore defaults with this command:
*RST
2.
Configure the desired waveform using one of the procedures previously
described for a sine, square, ramp, or arbitrary waveform.
3.
Configure the low jitter external trigger mode.
Examples
– The following commands configure the selected waveform to
be output following each pulse received on trigger link line 1. If a
subsequent trigger pulse is received while the waveform is being output,
the Model 6221 will immediately terminate the current wave and restart the
waveform. In between waveforms, the unit will output zero current.
4.
Arm and trigger the waveform to enable the output. The Model 6221 will
wait for an external trigger pulse on the specified trigger link line and start
the waveform within 1
μ
s:
5.
To stop the Model 6221 from responding to triggers and immediately stop
any waveform being generated by turning the source output off, send this
command:
SOUR:WAVE:EXTR:ENAB ON
‘ Turn on external trigger mode.
SOUR:WAVE:EXTR:ILIN 1
‘ Use TLINK Line 1 as input trigger.
SOUR:WAVE:EXTR:IGN OFF
‘ Restart wave on retrigger attempts.
SOUR:WAVE:EXTR:IVAL 0.00
‘ Output zero current between waves.
SOUR:WAVE:ARM
‘ Arm waveform.
SOUR:WAVE:INIT
‘ Enable output, wait on external triggers.
SOUR:WAVE:ABOR
‘ Stop generating waveforms.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Summary of Contents for 6220
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...