9-2
Limit Test and Digital I/O
Model 6220/6221 Reference Manual
Return to
Limit test
Overview
As shown in
, there is one limit test that can be performed on a DUT.
Limit 1 is used to test whether or not the Model 622x current source is in or out of
compliance. If the source is not in compliance (limit test PASS), all four lines of the
digital I/O port will be low. The Model 622x can be programmed to apply a specific
4-bit fail pattern on the digital I/O lines (
“Digital I/O connector,” on page 9-5
) if the
source is in compliance (limit test FAIL).
Figure 9-1
Limit 1 test (compliance)
Programming limit testing
The example below shows how to enable limit testing and set a failure pattern of 9
(1001). The FAIL? query is also used to determine if the limit test has passed or
failed.
With the above programming, a binary pattern of 1001 will appear on the digital
I/O lines if the limit test fails (source in compliance), and the FAIL? query will
return 1. A binary pattern of 0000 will appear on the digital I/O lines when the
source is not in compliance (limit test passes), and the FAIL? query will return a 0.
CALC3:LIM:STAT ON
‘ Enable limit test.
CALC3:LIM:SOUR2 9
‘ Set fail pattern to 9 (1001).
CALC3:LIM:FAIL?
‘ Query pass/fail of limit test.
Limit 1 Test
(Compliance)
Fail
Pass
(Not In Compliance)
(In Compliance)
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Summary of Contents for 6220
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...