Model 6220/6221 Reference Manual
Getting Started
1-25
Return to
To restore any setup
1.
Press
SETUP
to open the RESTORE SETUP menu:
2.
Select
USER
,
PRESET,
or
*RST:
•
USER
– Enter the desired value (0 to 4) and press
ENTER
.
•
PRESET
– Press
ENTER
to return to the PRESET defaults.
•
*RST
– Press
ENTER
to return to *RST defaults.
To select power-on setup
1.
Press the
SAVE
key to open the SAVED SETUP MENU.
2.
Select the
POWER ON
menu item.
3.
Select
BENCH
,
GPIB,
or
USER SETUP NUMBER
:
•
PRESET
– Press
ENTER
to select the PRESET defaults.
•
*RST
– Press
ENTER
to select the *RST defaults.
•
USER SETUP NUMBER
– Key in the desired user setup number and
then press
ENTER
.
Remote operation setups
Saving and restoring user setups
The *SAV and *RCL commands are used to save and recall user setups:
Restoring PRESet or *RST default setups
The SYSTem:PRESet command returns Model 622x to the PRESET defaults and
the *RST command returns it to the *RST defaults.
*SAV <NRf>
*RCL <NRf>
‘
‘
‘
Save present setup in memory.
Recall saved user setup from memory.
<NRf> = 0, 1, 2, 3, or 4
NOTE
User setups cannot be saved or recalled while Wave,
Sweep, Delta, Pulse Delta or Differential Conductance is
armed or running. The *SAV and *RCL commands will gen-
erate error +413 (Not allowed with mode armed).
A custom (list) sweep cannot be saved as a user setup.
Attempting to do so will generate error +528 Cannot save
CUSTOM sweep setup.
Source preset values are not aved aspart of a user setup.
SYSTem:PRESet
‘ Restore PRESET default setup.
*RST
‘ Restore *RST default setup.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Summary of Contents for 6220
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...