Source-Measure Concepts
5-3
Maximum compliance values
The maximum compliance values for the measurement ranges are summarized in Table 5-1.
Compliance examples
When the SourceMeter goes into real compliance, the “Cmpl” label for the compliance dis-
play will flash. When the SourceMeter goes into range compliance, the units label (“mA”) will
instead flash. For the following examples, labels in boldface indicate that they are flashing.
Measurement Range:
100mA
Compliance Setting:
Cmpl
: 075.000 mA
Flashing “Cmpl” indicates that
real compliance has occurred.
The output is clamped at 75mA.
Measurement Range:
10mA
Compliance Setting:
Cmpl: 075.000
mA
Flashing “mA” indicates that
range compliance has occurred.
The output is clamped at 10.5mA.
Table 5-1
Maximum compliance values
Measurement
range
Maximum compliance
value
200mV
2V
20V
200V
1pA*
10pA*
100pA*
1nA*
10nA*
100nA*
1µA
10µA
100µA
1mA
10mA
100mA
210mV
2.1V
21V
210V
1.05pA
10.5pA
105pA
1.05nA
10.5nA
105nA
1.05µA
10.5µA
105µA
1.05mA
10.5mA
105mA
* Only available when using the Remote PreAmp.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......