Source-Measure Concepts
5-21
Measure only (V or I)
Figure 5-11 shows the configurations for using the SourceMeter exclusively as a voltmeter
or ammeter. As shown in Figure 5-11A, the SourceMeter is configured to measure voltage-only
by setting it to source 0A and measure voltage.
CAUTION
V-Compliance must be set to a level that is higher than the measured volt-
age. Otherwise, excessive current will flow into the SourceMeter. This cur-
rent could damage the SourceMeter. Also, when connecting an external
voltage to the I-Source, set the output off state to the high-impedance
mode. See Section 12,
Output configuration
.
In Figure 5-11B, the SourceMeter is configured to measure current-only by setting it to
source 0V and measure current. Note that in order to obtain positive (+) readings, conventional
current must flow from IN/OUT HI to LO.
NOTE
If the Remote PreAmp is not used, use the INPUT/OUTPUT H1 and LO terminals on
the mainframe. Note however, that when not using the Remote PreAmp, the 100nA
through 1pA current ranges are not available.
6430
Sense Selection: 2-wire
I-Source
(0.00000pA
V-Meter
REMOTE
PreAmp
A. Measure Voltage Only
HI
LO
DUT
Voltage
Source
6430
Sense Selection: 2-wire
V-Source
(000.000mV)
I-Meter
REMOTE
PreAmp
B. Measure Current Only
HI
LO
DUT
Current
Source
Positive
Current
NOTE:
Positive current flowing out of input/output HI results
in positive (+) measurements.
IN/OUT
HIGH
IN/OUT
HIGH
Figure 5-11
Measure-only (V or I)
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......