Sweep Operation
9-11
SML 007
— Leakage Current Test
•
V, Measure I.
•
Limit 2 test – Min/max limits for current reading.
•
Summary – This test is the same as the test at memory location 004, except the source
voltage is reversed to properly bias the diode that was installed backwards.
Test results
— The test results for the 4-point source memory sweep are stored in the buffer.
Stored readings are accessed by pressing the RECALL key. Each of the four reading numbers
are preceded by a “P” or an “F” to indicate the “Pass” or “Fail” result of the corresponding test.
See Section 8 for details on the data store.
Con
fi
guring and running a sweep
Front panel sweep operation
Con
fi
guring a sweep
The sweep configuration menu is structured as follows and shown in Figure 9-7. Note that
bullets indicate the primary items of the sweep menu and dashes indicate the options of each
menu item. Using Section 1,
Rules to navigate menus
, go through the following menu to select
and configure the desired sweep.
CONFIGURE SWEEPS menu:
Press CONFIG then SWEEP to display the sweep configuration menu.
•
TYPE – Use this menu item to select the type of sweep:
-
STAIR – When the linear staircase sweep is selected, you will be prompted to enter
the START, STOP, and STEP levels.
-
LOG – When the logarithmic staircase sweep is selected, you will be prompted to
enter the START and STOP levels and specify the number of measurement points.
-
CUSTOM – With the custom sweep selected, you specify the number of measure-
ment points (# POINTS) in the sweep and the source level at each point (ADJUST
POINTS). With the INIT option, you can set a consecutive range of measurement
points in the sweep to a specific level. For example, assume that for a 20-point cus-
tom voltage sweep (# POINTS = 20), you want points 10 through 15 to be set for
1V. After selecting the INIT option, set the VALUE to +1.000000V, set the START
PT to 10, and set the STOP PT to 15.
-
SRC MEMORY – With the Source Memory Sweep selected, you specify the mem-
ory location START point to start the sweep (1 is the default) and the number of
memory location points (# POINTS) in the sweep. When configured to sweep past
point 100, the sweep automatically wraps around to point 1.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......