Remote Operations
Differences: remote vs. local operation
Operation enhancements (remote operation)
There are some source-measure operations you can do over the IEEE-488 bus and RS-232
interface that you cannot do from the front panel; these are summarized below.
Math expressions
There are
fi
ve math expressions available from the panel. All except the Percent Deviation
are available as pre-de
fi
ned math expressions for remote operation. However, remote operation
allows you to create up to
fi
ve user-de
fi
ned math expressions for a total of nine expressions. An
example program shows how to create Percent Deviation as a user-de
fi
ned math expression.
Concurrent measurements
With the use of the TOGGLE key, you can measure (display) two functions concurrently.
Using remote operation, you can perform concurrent measurements on all three functions (volt-
age, current, and resistance). See Section 17,
SENSe1 Subsystem
for details.
Local-to-remote transition
When changing from local to remote operation, the following actions occur:
•
The SourceMeter stops performing source-measure operations and returns to the idle
state (ARM annunciator off).
•
All sweep operations are aborted.
•
All menus are exited.
•
All pending front panel commands are aborted.
•
Source and compliance editing are disabled.
•
Data in the sample buffer is lost (i.e., :FETCh?, :CALC1:DATA?, and :CALC2:DATA?
will not return any data until readings are taken while in remote).
•
Concurrent measurements are enabled.
•
All other settings are not affected, including those for the :TRACe buffer (data store).
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......