15-6
Common Commands
*TRG programming example
The command sequence in Table 15-4 con
fi
gures the SourceMeter to be controlled by bus
triggers. The last command, which sends a bus trigger, triggers one measurement. Each subse-
quent bus trigger will also trigger a single measurement.
NOTE
With :ARM:SOURce BUS selected, do not send any commands (except *TRG, GET,
DCL, SDC, IFC, and ABORt) while performing source-measure operations. Other-
wise, erratic operation will occur.
*TST?
—
self-test query
Run self test and read result
Use this query command to perform a checksum test on ROM. The command places the
coded result (0 or 1) in the Output Queue. When the SourceMeter is addressed to talk, the
coded result is sent from the Output Queue to the computer.
A returned value of zero (0) indicates that the test passed, and a value of one (1) indicates
that the test failed.
*WAI
—
wait-to-continue
Wait until previous commands are completed
Effectively, the *WAI command is a No-Op (no operation) for the SourceMeter and thus,
does not need to be used.
Two types of device commands exist:
•
Sequential commands
—
A command whose operations are allowed to
fi
nish before the
next command is executed.
•
Overlapped commands
—
A command that allows the execution of subsequent com-
mands while device operations of the Overlapped command are still in progress.
The *WAI command is used to suspend the execution of subsequent commands until the
device operations of all previous Overlapped commands are
fi
nished. The *WAI command is
not needed for Sequential commands.
Table 15-4
*TRG programming example
Command
Description
*RST
:ARM:SOUR BUS
:ARM:COUN INF
:OUTP ON
:INIT
*TRG
Restore GPIB defaults.
Select BUS control source.
Set arm layer count to in
fi
nite.
Turn on output.
Take SourceMeter out of idle.
Trigger one measurement.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......