SCPI Command Reference
17-27
Assign unit suf
fi
x
:UNITs <name>
:CALCulate[1]:MATH:UNITs <name>
Specify units for user-de
fi
ned calculation
Parameters
<name> = Three ASCII characters enclosed in single or double quotes
Query
:UNITs?
Query units for user-de
fi
ned calculation
Description
This command is used to specify the units suf
fi
x name for a user-de
fi
ned
math calculation. Use three ASCII characters for the units suf
fi
x name. If
using fewer than three characters, add spaces to the right of the units name
in the string. For example, if the units name is
“
Z
”
, send it as follows:
:calc:math:unit
“
Z
”
The units name can also be enclosed in single quotes as follows:
:calc:math:unit
‘
Z
’
De
fi
ne math expression
[:EXPRession] <form> or [:DEFine] <form>
:CALCulate[1]:MATH[:EXPRession] <form>
De
fi
ne math formula
:CALCulate[1]:MATH[:EXPRession][:DEFine] <form>
De
fi
ne math formula
Parameters
<form> = mathematical formula using instrument readings, numbers and
standard math operator symbols. See
Description
for details.
Query
:MATH?
Query user-de
fi
ned math expression
Description
Use either of these two commands to de
fi
ne a math formula using measure
and source readings, numeric constants, and standard math operator sym-
bols. After the math expression is de
fi
ned, it will be assigned to the name
that was created using the :NAME command and will become the selected
math expression. See :NAME for more details.
Valid parameter names for measure and source readings include:
VOLTage
Use V-Measure or V-Source reading
CURRent
Use I-Measure or I-Source reading
RESistance
Use ohms reading
TIME
Use timestamp reading
Valid math operators and their operations are listed as follows:
+
Add
-
Subtract
*
Multiply
/
Divide
^
Exponent
log
Logarithmic, base 10
ln
Natural log
sin
Sine
cos
Cosine
tan
Tangent
exp
e
x
abs
Absolute value
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......