1-6
Getting Started
Product overview
The SourceMeter combines a precise, low-noise, highly stable DC power supply with a low-
noise, highly repeatable, high-impedance multimeter and a remote preamplifier for ultra low
current measurements. It has 0.012% basic accuracy with 5
½
-digit resolution. At 5
½
digits, the
SourceMeter delivers 520 readings/second over the IEEE-488 bus. At 4
½
digits, it can read up
to 2000 readings/second into its internal buffer. The unit has broad source and measurement
ranges:
•
Source voltage from 5µV to 210V; measure voltage from 1µV to 211V.
•
Source current from 0.5fA to 105mA; measure current from 10aA to 105.5mA.
•
Measure resistance from 100µ
Ω
(<100µ
Ω
in manual ohms) to 21.1T
Ω
.
•
Maximum source power is 2.2W.
NOTE
The Model 6430 is Y2K compliant.
Some additional capabilities of the SourceMeter include:
•
Perform measurements at the DUT using the small Remote PreAmp.
•
Concurrent measurements of all three functions over the remote interface.
•
Source-measure sweep capabilities (linear and logarithmic staircase sweeps, source
sweep list of up to 1000 points, memory sweep of up to 100 instrument setups).
•
6-wire ohms measurement with programmable I-source or V-source with V or I clamp.
•
2.2W, 4-quadrant source and sink operation.
•
Up to 11 stages of limit testing with a built-in comparator for pass/fail testing.
•
Digital I/O for stand-alone binning operations or interface to component handler.
•
Programming language and remote interfaces — The SourceMeter uses the SCPI pro-
gramming language and two remote interface ports (IEEE-488/GPIB and RS-232C).
•
Trigger-Link interface to Keithley Series 7000 switching hardware.
•
Math expressions — 5 built-in, up to 5 user-defined (bus only).
•
Reading and setup storage — Up to 2500 readings and seven setups (five user defaults,
factory default, *RST default) can be stored and recalled.
•
Closed-cover calibration — The instrument can be calibrated either from the front
panel or remote interface.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......