Basic Source-Measure Operation
Measure only
Front panel measure only
In addition to being used for conventional source-measure operations, the SourceMeter can
also be used to measure only voltage or current. Perform the following steps to use the
SourceMeter to measure voltage or current:
1.
Select source-measure functions.
Measure voltage only (voltmeter)
— Press SOURCE I to select the I-Source, and
press MEAS V to select the voltage measurement function.
Measure current only (ammeter)
— Press SOURCE V to select the V-Source, and
press MEAS I to select the current measurement function.
2.
Set source and compliance levels.
Use the editing procedure provided in step 2 of
Basic source-measure procedure
to edit
the source and compliance levels as follows:
a.
Select the lowest source range and set the source level to zero (0.00000pA or
000.000mV).
b.
Set compliance to a level that is higher than the expected measurement.
C
A
U
T
I
O
N
When using the SourceMeter as a voltmeter, V-Compliance must be set
higher than the voltage that is being measured. Failure to do this could
result in instrument damage due to excessive current that will flow into the
SourceMeter.
3.
Select range.
Use the RANGE
▲
and
▼
keys to select a fixed measurement range that will accom-
modate the expected reading. Use the lowest possible range for best accuracy.
When measuring current, AUTO range can be used instead. The SourceMeter will auto-
matically go to the most sensitive range. When measuring voltage, DO NOT use AUTO
range. See the following CAUTION.
C
A
U
T
I
O
N
When using the SourceMeter as a voltmeter only, DO NOT use AUTO
range and NEVER select a measurement range that is below the applied
signal level. For these conditions, high current will be drawn from the
external source. This high current could damage the external source or test
circuit.
4.
Connect voltage or current to be measured. Connect the DUT to the SourceMeter using
2-wire connections. See Figure 2-2.
5.
Turn output on. Press the ON/OFF key to turn the output on.
6.
Take a reading from the display.
7.
When finished, turn output off.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......