Triggering
10-7
Bench defaults
The bench defaults are listed as follows. They are also denoted in Figure 10-1 by the “
✛
”
symbol.
•
Arm-In Event = Immediate
•
Trigger-In Source = Immediate
•
Arm Count = 1
•
Trigger Count = 1
•
Trigger Delay = 0.0 sec
•
Delay Action = 0.001 sec
•
Source Trigger In Event = On
•
Delay Trigger In Event = Off
•
Measure Trigger In Event = Off
•
Trigger Out Events = All Trigger Out Events are disabled (off)
•
Arm Out Events = Off
•
Event Detection Bypasses = Never (both layers)
When the output is turned ON, the SourceMeter will run in a continuous loop around the
trigger model. After each Measure Action, operation will continue at the top of the trigger
model. The SourceMeter can be returned to idle by turning the output OFF.
Operation summary
The trigger model is designed to offer versatility for the various source-measure applica-
tions. Typically, it allows you to perform a specified number of measurements at various source
levels.
For example, assume you want to perform three measurements each at two different
V-source levels (1V and 2V). To do this, set the arm count to two, the trigger count to three,
and use a 6-point Custom Sweep configured as follows:
P0000 = 1V
P0001 = 1V
P0002 = 1V
P0003 = 2V
P0004 = 2V
P0005 = 2V
When the sweep is started, operation falls into the Trigger Layer and performs three mea-
surements at the 1V source level. Operation then loops back into the Trigger Layer to perform
three measurements at the 2V source level. The six readings are stored in the buffer.
Note that after the sweep is finished, the SourceMeter does not return to idle. Operation con-
tinues at the top of the trigger model. Subsequent measurements are performed at the 2V level
and are not stored in the buffer.
For details on the Custom Sweep, see Section 9.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......