12-10
Digital I/O Port, Interlock, and Output Configuration
Output con
fi
guration programming example
Table 12-3 lists the command sequence for output con
fi
guration. These commands set up the
SourceMeter as follows:
•
Interlock: enabled
•
Output-off mode: normal
•
Auto-off mode: on
NOTE
Connect pins 8 and 9 of the Digital I/O Port together to simulate a closed interlock
switch. Otherwise, the unit will not turn on its output when the measurement is made.
Table 12-3
Output configuration programming example
Command
Description
*RST
:SOUR:VOLT 10
:OUTP:INT:STAT ON
:OUTP:SMOD NORM
:SOUR:CLE:AUTO ON
:READ?
Restore GPIB defaults.
Output 10V.
Enable interlock.*
Select normal output-off mode.
Enable auto-off mode.
Trigger and acquire readings.
*Connect pins 8 and 9 of Digital I/O Port to simulate closed interlock switch.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......