17-64
SCPI Command Reference
SOURce subsystem
This subsystem is used to con
fi
gure and control the I-Source and V-Source, and to set the
logic level (high or low) of each digital output line. The commands for this subsystem are sum-
marized in Table 17-6.
SOURce[1]
Use the following commands to con
fi
gure and control the I-Source and V-Source. At the end
of this subsystem are program examples of sweeps and lists.
Control source output-off
[:IMMediate]
:SOURce[1]:CLEar[:IMMediate]
Turn source output off
Description
This command is used to turn off the source output. The output will turn off
after all programmed source-measure operations are completed and the
instrument returns to the idle state.
Note that if auto output-off is enabled, the source output will automatically
turn off. See the next command.
:AUTO
:SOURce[1]:CLEar:AUTO <b>
Control auto output-off
Parameters
<b> =
1 or ON
Enable auto output-off
0 or OFF
Disable auto output-off
Query
:AUTO?
Query state of auto output-off
Description
This command is used to control auto output-off for the source. With auto
output-off enabled, an :INITiate (or :READ? or MEASure?) will start
source-measure operation. The output will turn on at the beginning of each
SDM (source-delay-measure) cycle and turn off after each measurement is
completed.
With auto output-off disabled, the source output must be on before an
:INITiate or :READ? can be used to start source-measure operation. The
:MEASure? command will automatically turn on the source output. Once
operation is started, the source output will stay on even after the instrument
returns to the idle state. Auto output-off is the *RST and :SYSTem:PRESet
default.
WARNING
With auto output-off disabled, the source output will remain on after all
programmed source-measure operations are completed. Beware of hazard-
ous voltage that may be present on the output terminals.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......