Performance Verification
18-5
Test resistor construction
The 1T
Ω
and 10T
Ω
test resistors used to test the 2T
Ω
and 20T
Ω
resistance ranges should be
mounted in a shielded, guarded enclosure like the one shown in Figure 18-1. The resistors
should be properly characterized to the accuracy stated in Table 18-1 before use. Connect the
resistors terminals to the triax jack center conductor and outer shell. Connect the guard shield
to the triax jack inner ring.
WARNING
To avoid a shock hazard, do not connect guard to an exposed shield.
Veri
fi
cation limits
Most of the veri
fi
cation reading limits stated in this section have been calculated using only
the Model 6430 one-year accuracy speci
fi
cations, and most reading limits do not include test
equipment uncertainty. (However, 1pA-100nA and 2G
Ω
-200G
Ω
range limits do include the
uncertainty of the Model 5156 Electrometer Calibration Standard.) If a particular measurement
falls outside the allowable range, recalculate new limits based both on Model 6430 speci
fi
ca-
tions and corresponding test equipment speci
fi
cations.
Example limits calculation
As an example of how veri
fi
cation limits are calculated, assume you are testing the 20V DC
source range using a 20V output value. Using the Model 6430 20V range one-year accuracy
speci
fi
cation of ±(0.02% of 2.4mV offset), the calculated output limits are:
Output limits = 20V ± [(20V
×
0.02%) + 2.4mV]
Output limits = 20V ± (0.004 + 0.0024)
Output limits = 20V ± 0.0064V
Output limits = 19.9936V to 20.0064V
Center
Conductor
Triax Jack
Inner Ring
Shell
Guard
Shield
1T
Ω
or 10T
Ω
Resistor
Figure 18-1
Test resistor
construction
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......