Performance Verification
18-19
5.
Verify output voltage accuracy for each of the voltages listed in Table 18-8. For each
test point:
•
Select the correct source range.
•
Set the Model 6430 output voltage to the indicated value.
•
Verify that the multimeter reading is within the limits given in the table.
6.
Repeat the procedure for negative output voltages with the same magnitudes as those
listed in Table 18-8.
Remote PreAmp voltage measurement accuracy
Follow the steps below to verify that Model 6430 Remote PreAmp voltage measurement
accuracy is within speci
fi
ed limits. The test involves setting the source voltage to full-range val-
ues, as measured by a precision digital multimeter, and then verifying that the Model 6430 volt-
age readings are within required limits.
1.
With the power off, connect the digital multimeter to the PreAmp IN/OUT HIGH jack
as shown in Figure 18-5. (Connect the IN/OUT HIGH jack to the DMM INPUT HI and
LO jacks using the adapters and coax cable as shown.)
2.
Select the SourceMeter 10µA measurement range.
3.
Select the multimeter DC volts function.
4.
Set the Model 6430 to both source and measure voltage by pressing the SOURCE V
and MEAS V keys, and make sure the source output is turned on.
5.
Verify output voltage accuracy for each of the voltages listed in Table 18-9. For each
test point:
•
Select the correct source range.
•
Set the Model 6430 output voltage to the indicated value as measured by the digital
multimeter.
•
Verify that the Model 6430 voltage reading is within the limits given in the table.
NOTE
It may not be possible to set the voltage source to the specified value. Use the closest
possible setting, and modify reading limits accordingly.
Table 18-8
Remote PreAmp output voltage accuracy limits
Model 6430
source range
Model 6430
output voltage setting
Output voltage limits
(1 year, 18°-28°C)
200mV
2V
20V
200V
200.000mV
2.00000V
20.0000V
200.000V
199.360 to 200.640mV
1.99900 to 2.00100V
19.9936 to 20.0064V
199.936 to 200.064V
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......