18-26
Performance Verification
Remote PreAmp resistance measurement accuracy
Use the following steps to verify that Model 6430 Remote PreAmp resistance measurement
accuracy is within speci
fi
ed limits. This procedure involves applying accurate resistances from
a resistance calibrator or standard resistor and then verifying that Model 6430 resistance mea-
surements are within required limits.
NOTE
Before making resistance measurements, put the Model 6430 in the AUTO SOURCE
mode (press CONFIG, then
Ω
, select SOURCE, select AUTO, then press ENTER).
CAUTION
Before testing the 20
Ω
and 200
Ω
ohms ranges, make sure your resistance
calibrator can safely handle the default test currents for those ranges
(100mA and 10mA for the 20
Ω
and 200
Ω
ranges respectively). If not, use
the Model 6430 MANUAL ohms mode, and set the test current to the max-
imum safe calibrator current. Note that Model 6430 measurement accu-
racy is reduced and reading limits should be recalculated using the
additional uncertainty when using MANUAL ohms. See the specifications
in Appendix A for details.
20
Ω
-200M
Ω
range accuracy
1.
With the power off, connect the Remote PreAmp IN/OUT HIGH and SENSE jacks to
the resistance calibrator as shown in Figure 18-8. (Connect PreAmp IN/OUT HIGH to
calibrator OUTPUT; PreAmp SENSE to calibrator SENSE using the adapters and
cables shown.)
2.
Select the resistance calibrator external sense mode.
3.
Press MEAS
Ω
to select the ohms measurement function, and make sure the source out-
put is turned on.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......