19-14
Calibration
Remote PreAmp calibration
Use the procedures discussed below to calibrate the Remote PreAmp together with the
mainframe.
NOTE
The mainframe must be separately calibrated before calibrating the Remote
PreAmp. See “Mainframe calibration” earlier in this section for information on cal-
ibrating the mainframe.
Connecting Remote PreAmp to the mainframe
WARNING
Potentially hazardous source voltage is routed from the mainframe to the
Remote PreAmp via the PreAmp cable. Adhere to the following safety pre-
cautions to prevent electric shock:
•
The SourceMeter must be turned off before connecting (or disconnect-
ing) the Remote PreAmp to the mainframe.
•
When not using the Remote PreAmp, disconnect the PreAmp cable at
the PreAmp connector on the mainframe. DO NOT leave the PreAmp
cable connected to the mainframe if the other end is not connected to
the Remote PreAmp.
•
ALWAYS re-install the plastic safety cover onto the mainframe
PreAmp connector whenever the Remote PreAmp is not being used.
Use the supplied PreAmp cable to connect the Remote PreAmp to the mainframe as follows:
1.
From the front panel of the SourceMeter, turn the POWER off.
2.
Connect the PreAmp cable to the Remote PreAmp. The PreAmp connector on the
Remote PreAmp is labeled
“
Mainframe.
”
3.
At the rear panel of the mainframe, remove the plastic safety cover from the PreAmp
connector. This PreAmp connector is labeled
“
REMOTE PreAmp.
”
The plastic cover is
secured to the connector with two screws. Hold on to the plastic cover and the retaining
screws. Whenever the Remote PreAmp is not being used, the plastic safety cover must
be re-installed on the mainframe PreAmp connector.
4.
Connect the other end of the PreAmp cable to the mainframe.
Remote PreAmp calibration menu
When the Remote PreAmp is connected to the mainframe, you can access the PreAmp cali-
bration menu by pressing the MENU key and selecting CAL. Menu choices include:
•
UNLOCK: Enables Remote PreAmp calibration.
•
V-BURDEN: Performs voltage burden calibration.
•
GAIN: Performs gain calibration.
•
OFFSET: Performs offset calibration.
•
LOCK: Locks out Remote PreAmp calibration.
•
CHANGE-PASSWORD: Allows you to change the calibration password.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......