C-2
Data Flow
Introduction
Data
fl
ow for remote operation is summarized by the block diagram shown in Figure C-1.
Refer to this block diagram for the following discussion.
The SENSE block represents the basic measured readings of voltage, current and resistance.
If Filter is enabled, the readings will be
fi
ltered. The SENSE block also measures time for the
timestamp.
When the INITiate command is sent, the programmed number of source-measure operations
are performed and the respective data is temporarily stored in the Sample Buffer. For example,
if 20 source-measure operations were performed, then 20 sets of data will be stored in the Sam-
ple Buffer. Data from this buffer is then routed to other enabled data
fl
ow blocks.
SENSE
(Measurements)
Volts, Amps, Ohms,
Timestamp, Filter
Sample
Buffer
CALC1
Math
Expression
CALC2
Limit Tests
NULL (Rel)
Trace
(Data Store)
FETCh?
READ?
MEAS? CALC1:DATA?
CALC2:DATA?
CALC3:DATA?
TRACe:DATA?
CALC3
Min, Max, Sdev
Mean, Pk-Pk
Figure C-1
Data flow block
diagram
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......