C-4
Data Flow
TRACe:DATA?
If the data store is enabled, Sample Buffer data, CALC1 results, and CALC2 readings
become available to the TRACE block for storage. The selected feed determines which group
of readings are stored.
The TRACe:DATA? command reads the entire contents of the data store.
CALCulate3:DATA?
Statistical information (minimum, maximum, mean, standard deviation, and peak-to-peak)
is available for measure readings stored in the buffer. If the readings in the data store came
directly from the Sample Buffer, then the selected statistic calculation will be performed on all
enabled measurement functions. The calculation results are returned in the following
fi
xed
order:
VOLTage result, CURRent result, RESistance result
When the TRACE buffer is feeding off CALC1 or CALC2, the selected statistic calculation
is performed on the selected feed. Thus, a single statistic result is acquired for each reading
stored in the data store.
The CALCulate3:DATA? command acquires the results of the selected calculation.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......