Getting Started
1-25
Editing source and compliance values
Use the following keys to edit source and compliance values:
•
EDIT: selects the source or compliance display field for editing. A blinking cursor will
appear in the field to be edited. If no key is pressed within a few seconds, the edit mode
will be cancelled automatically.
•
EDIT and : places the display cursor on the display digit to be changed.
•
SOURCE
▲
or
▼
: increments or decrements the source or compliance value. Note that
pressing either of these keys will automatically enable the source edit mode.
•
RANGE
▲
or
▼
: selects the source or compliance range.
•
Numeric keys (0-9): allow you to directly enter source or compliance values.
•
EXIT: exits the edit mode without waiting for the time-out period.
The basic procedure for editing source and compliance values is outlined below. See
Basic
source-measure procedure
in Section 3 for more details.
1.
Press the EDIT key until the blinking cursor is in either the source or compliance dis-
play field to be edited.
2.
If desired, use the RANGE
▲
and
▼
keys to select the desired source or compliance
range.
3.
To simply increment or decrement the display value, use the EDIT and keys to
place the blinking cursor on the digit to be changed, then increment or decrement the
value with the SOURCE
▲
and
▼
keys. Note that the source or compliance value will
be updated immediately; you need not press ENTER to complete the process.
4.
To enter the source or compliance value directly, simply key in the desired value with
the numeric keys while the cursor is blinking. Again, the source or compliance value
will be updated immediately.
Toggling the source and measure display
fi
elds
Normally the measured reading value will appear in the upper, main display line, while the
source and compliance values will appear in the left and right fields respectively of the lower
display line. You can toggle the source and measure display fields by pressing the TOGGLE
key to place the source and measure values in the desired positions.
▲
▲
▲
▲
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......