Basic Source-Measure Operation
Basic circuit configuration
The fundamental source-measure configuration for the SourceMeter (with Remote PreAmp)
is shown in Figure 3-1, where the Source is either the V-Source or the I-Source. If not using the
Remote PreAmp, Input/Output HI and LO is accessed at the rear panel of the mainframe.
N
OT
E
When using the Remote PreAmp, nothing should be connected to INPUT/OUTPUT
HI banana jack on the mainframe. If using mainframe INPUT/OUTUT HI, the cable
to the Remote PreAmp should be disconnected.
Operation considerations
The following paragraphs discuss warm-up period, auto zero, V-source protection, and
source delay.
Warm-up
The SourceMeter must be turned on and allowed to warm up for at least one hour to achieve
rated accuracies. See Appendix A for specifications.
Auto zero
Every A/D conversion (reading) is calculated from a series of zero, reference, and signal
measurements. With auto zero enabled, all three of these measurements are performed for each
reading to achieve rated accuracy. With auto-zero disabled, zero and reference are not mea-
sured. This increases measurement speed, but zero drift will eventually corrupt accuracy. With
auto zero disabled, periodically change measurement speed.
Mainframe
I-Meter
Source
V-Meter
REMOTE
PreAmp
HI
LO
A
IN/OUT
HIGH
i
g
u
r
e 3
-
1
undamental
ource-measure
onfiguration
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......