6482-901-01 Rev. A / August 2012
Return to
A-5
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Appendix A: Status and Error Messages
Number
Error message
Event
1
Status register
2
Bit
+000
No error
SE
+100
+101
+102
+103
+104
+105
+106
+107
+108
+109
+111
+114
Measurement events:
Limit 1 failed
Low limit 2 failed
High limit 2 failed
Low limit 3 failed
High limit 3 failed
Active limit tests passed
Reading available
Reading overflow
Buffer available
Buffer full
OUTPUT enable asserted
Source in compliance
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
SE
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
Measurement Event
0
1
2
3
4
5
6
7
8
9
11
14
+200
Standard events:
Operation complete
SE
Standard Event
0
+300
+303
+305
+306
+310
Operation events:
Device calibrating
Device sweeping
Waiting in trigger layer
Waiting in arm layer
Entering idle layer
SE
SE
SE
SE
SE
Operation Event
Operation Event
Operation Event
Operation Event
Operation Event
0
3
5
6
10
+408
+414
Questionable events:
Questionable Calibration
Command Warning
SE
SE
Questionable Event
Questionable Event
8
14
Table A-1 (cont.)
Status and error messages
1
EE = Error Event, SE = Status Event, SYS = System Error Event
2
Use following queries to read status registers:
Standard Event: *ESR?
Operation Event: STAT:OPER?
Measurement Event: STAT:MEAS?
Questionable Event: STAT:QUES?
Summary of Contents for 6482
Page 3: ......