6482-901-01 Rev. A / August 2012
Return to
17-29
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Section 17: Calibration
Figure 17-12
Channel 2 current calibration connections (20
μ
A to 20 mA ranges)
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
SERIAL NO.
120
LINE RATING
50, 60Hz
60 VA MAX
FUSE
LINE
630 mAT
(SB)
315 mAT
(SB)
100 VAC
120 VAC
220 VAC
240 VAC
CAT
I
MADE IN
U.S.A.
OUTPUT ENABLE
RS-232
TRIGGER LINK
INPUT
RATINGS MAX.
30 V @ 20mA
CHANNEL 1
CHANNEL 2
RATINGS MAX.
COMMON
MODE
200V
!
!
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
ANALOG
OUT
RATINGS MAX.
30 V @ 20mA
VOLTAGE
SOURCE
OUT
30 V @ 20mA
Model 6482
Model 4801
low noise
coaxial cable
HI
LO
Fluke Model 5700A multifunction calibrator
Model 7078-TRX-BNC
3-slot male triax to
BNC adapter
Model BG-18
Dual banana to
BNC coaxial
adapter. Connect
output LO to shield.
Output
V
Ω
A
Channel 2
INPUT
Summary of Contents for 6482