Limit Tests
8-3
The 2-stage limit testing process is shown in
. If limit 1 fails, the L1 message is
displayed and the test is finished. Limit 2 is not tested because the pass band relationship
between the two stages implies that if limit 1 fails, limit 2 must also fail. If limit 1 passes,
the limit 2 test is performed. If limit 2 fails, the L2 message is displayed. If both limit 1
and limit 2 pass, the OK message is displayed. The display messages for limit tests are
summarized in
.
A test is only performed if it is enabled. Therefore, you can perform a single-stage test or a
2-stage test. In the flowchart (
), operation simply proceeds through a disabled
test.
Table 8-1
Test limit display messages
Display
Message
Limit 1
Test Result
Limit 2
Test Result
:OK
Pass
Pass
:L1
Fail
Not Performed
:L2
Pass
Fail
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......