10-2
Status Structure
Model 6485 Picoammeter Instruction Manual
Overview
Model 6485 provides a series of status registers and queues allowing the operator to mon-
itor and manipulate the various instrument events. The status structure is shown in
. The heart of the status structure is the status byte register. This register can be
read by the users test program to determine if a service request (SRQ) has occurred and
what event caused it.
Status byte and SRQ
— The status byte register receives the summary bits of four status
register sets and two queues. The register sets and queues monitor the various instrument
events. When an enabled event occurs, it sets a summary bit in the status byte register.
When a summary bit of the status byte is set and its corresponding enable bit is set (as pro-
grammed by the user), the RQS/MSS bit will set to indicate that an SRQ has occurred.
Status register sets
— A typical status register set is made up of a condition register, an
event register, and an event enable register. A condition register is a read-only register that
constantly updates to reflect the present operating conditions of the instrument.
When an event occurs, the appropriate event register bit sets to 1. The bit remains latched
to 1 until the register is reset. When an event register bit is set and its corresponding enable
bit is set (as programmed by the user), the output (summary) of the register will set to 1,
which in turn sets the summary bit of the status byte register.
Queues
— Model 6485 uses an output queue and an error queue. The response messages
to query commands are placed in the output queue. As various programming errors and
status messages occur, they are placed in the error queue. When a queue contains data, it
sets the appropriate summary bit of the status byte register.
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......