Model 6485 Instruction Manual
Calibration
16-7
5.
Enter the current calibration code on the display. (Factory default: 006485.) Use
the up and down RANGE keys to select the letter or number, and use the left and
right arrow keys to choose the position. Press ENTER to complete the process, and
the unit will display:
CAL UNLOCKED
Followed by:
NEW CODE? N
6.
With N displayed, press ENTER.
Offset voltage calibration
Before performing the remaining calibration steps, perform input bias current and offset
voltage calibration as outlined below.
1.
From the calibration menu, use the up or down RANGE key to display the follow-
ing:
CAL: VOFFSET
2.
Press ENTER. The instrument will prompt as follows:
INPUT CAP
3.
Connect the BNC shielding cap to the rear panel INPUT jack.
4.
Press ENTER to complete offset voltage calibration.
5.
Press EXIT to return to normal display.
Current calibration
20
µ
A-20mA range calibration
1.
Connect the BNC shielding cap to the Model 6485 rear panel INPUT jack.
2.
Select the Model 6485 20
µ
A range.
3.
Press MENU, select CAL, then press ENTER. At the CAL: RUN prompt, press
ENTER. The unit will prompt for the zero calibration point:
20
µ
A ZERO
4.
With the BNC shielding cap connected to the INPUT jack for a 0
µ
A input current,
press ENTER.
5.
The unit will prompt for the positive full-scale cal point:
+20
µ
A CAL
6.
Connect the current calibrator to the Model 6485 INPUT jack, as shown in
7.
Press ENTER. The unit will prompt for the positive full-scale current:
+20.00000
µ
A
8.
Set the calibrator current to +20.00000
µ
A then adjust the display to agree with the
calibrator current.
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......