D-10
DDC Emulation Commands
Model 6485 Picoammeter Instruction Manual
Figure D-1
U0 Status word
TERMINATOR
: CR LF
= LF CR
, LF
. CR
None
PREFIX
ZERO CHECK
0 = Off
1 = On
RANGE
0 = Auto On
1 = 2nA
2 = 20nA
3 = 200nA
4 = 2
m
A
5 = 20
m
A
6 = 200
m
A
7 = 2mA
8 = 20mA
9= Auto Off
LOG
0 = Off
1 = On
RELATIVE
0 = Off
1 = On
EOI
Value
Send EOI
Bus hold-off on X
0
Yes
Enable
1
No
Enable
2
Yes
Disable
3
No
Disable
SRQ Error Mask*
00 = SRQ Disabled
01 = IDDCO
02 = IDDC
03 = IDCCO or IDCC
04 = Not in Remote
05 = Not in Remote or IDDCO
06 = Not in Remote or IDDC
07 = Not in Remote, IDDCO,
or IDDC
TRIGGER
0 = Continuous on Talk
1 = One-shot on Talk
2 = Continuous on Get
3 = One-shot on Get
4 = Continuous on X
5 = One-shot on X
6 = Continuous on External Trigger
7 = One-shot on External Trigger
*Also see Byte 5 in Status Byte
4
8
5
C
D
R
Z
K
T
M
d
M
e
Y
STATUS WORD
SRQ Data Mask* (see Legend)
00 = SRQ Disabled
01 = Reading Overflow
02 = DSF
03 = DSF or Overflow
08 = Done
09= Done or Overflow
10 = DSF or Done
11 = DSF or Done or Overflow
16 = Busy
17 = Busy or Overflow
18 = DSF or Busy
19= DSF or Busy or Overflow
24 = Busy or Done
25 = Busy, Done, or Overflow
26 = DSF or Busy or Done 27 = DSF or Busy or Done or Overflow
SRQ Data Mask Legend
Overflow = Reading Overflow
DSF
= Data Store Full
Done
= Reading Done
Busy
= Busy
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......