Model 6485 Picoammeter Instruction Manual
Applications Guide
I-3
voltage difference is equal to the voltage burden of the picoammeter. Since the top and the
bottom of the insulator are at nearly the same potential, no significant current will flow
through it, and nearly all the current from the device under test will flow through the
picoammeter as well.
Figure I-1
Guarding to reduce leakage currents
Input bias current
An ideal picoammeter would read 0A with an open input. In practice, however, ammeters
do have some current that flows when the input is open. This current is known as the input
bias (offset) current and may be large enough to corrupt low current measurements. The
input bias current for Model 6485 is included in the offset portion of the accuracy specifi-
cation.
Voltage burden
The input resistance of the picoammeter causes a small voltage drop across the input ter-
minals. This voltage is known as the voltage burden. If the voltage burden is large in rela-
tion to the voltage of the measured circuit, then significant measurement errors will occur.
Refer to
to see how voltage burden affects current measurements. Assume V
S
is
5mV and R
S
is 5k
Ω
to configure a 1
µ
A current source (5mV/5k
Ω
= 1µA). An ideal
picoammeter with zero voltage burden would allow 1
µ
A to flow and measure it accurately.
In practice however, every picoammeter has a voltage burden. If the voltage burden (V
B
) is
1mV, the current that flows will be:
Programmable
V-Source
+
-
6485
Picoammeter
Metal
Shield
Measured
Current
HI
LO
HI
LO
Equivalent Circuit
DUT
Test Fixture
Insulators
A
I
M
V
S
V
B
–
R
S
--------------------
5mV
1mV
–
5k
Ω
-------------------------------
0.8
µ
A
=
=
=
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......