I-18
Applications Guide
Model 6485 Picoammeter Instruction Manual
Applications
The following applications require an external voltage source.
NOTE
External triggering and delay are covered in
.
With the proper use of external triggering between Models 6485 and voltage source, the
tests can be automated. All of the applications require a bias time or delay, which can be
provided by the delay feature of Model 6485. When Model 6485 is triggered, a measure-
ment will not be performed until the delay period expires.
Diode leakage current
shows how to measure the leakage current for a diode. By sourcing a positive
voltage, the leakage current through the diode will be measured. Note that if you source a
negative voltage, you will forward bias the diode. Resistor R is used to limit current in the
event that the diode shorts out or it becomes forward biased. Select a value of R that will
limit current to 20mA or less.
A profile for leakage current can be developed by measuring current at various voltage
levels. For example, program the voltage source to source from 1 to 10V in 1V steps. With
the proper use of external triggering, the Model 6485 performs a current measurement on
each voltage step. To ensure that the voltage is settled before each current measurement,
you can program Model 6485 for a delay. For example, if you program Model 6485 for a
one second delay, each measurement will be performed after the voltage step is allowed to
settle for one second. The current measurements can be stored in the buffer.
NOTE
Buffer operation is covered in
.
Figure I-14
Connections; diode leakage current test
A
+
-
Programmable
V-Source
6485
Picoammeter
Metal
Shield
HI
LO
HI
LO
R
Diode
Equivalent Circuit
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......