1-14
Getting Started
Model 6485 Picoammeter Instruction Manual
Power-up sequence
The following power-up sequence occurs when the Model 6485 is turned on:
1.
The Model 6485 performs self-tests on its EPROM and RAM with all digits and
annunciators turned on. If a failure is detected, the instrument momentarily dis-
plays an error message and the ERR annunciator turns on. Error messages are
listed in
NOTE
If a problem develops while the instrument is under warranty, return it to
Keithley Instruments Inc., for repair.
2.
If the instrument passes the self-tests, the firmware revision levels are displayed.
For example:
6485 B01
3.
After the firmware revision levels are displayed, the detected line frequency is dis-
played.
For example:
FREQ: 60Hz
4.
After the detected line frequency is displayed, information on the selected remote
interface is displayed:
a.
GPIB
— If the GPIB is the selected interface, the instrument will display the
selected language (SCPI or DDC) and primary address.
Examples:
SCPI ADDR: 14
DDC ADDR: 14
b.
RS-232
— If RS-232 is the selected interface, the instrument will display the baud
rate setting. For example:
RS-232: 9600b
5.
If the FACTory setup is selected as the power on setup, the unit is placed in the
default reading mode after the communication information is displayed. If a setup
other than FACTory is selected, the configured setup will be displayed. For exam-
ple, if the USR1 setup (User Setup #1) is selected:
USING USR1
NOTE
To configure power-on set up:
-Display PWR-ON: menu (press CONFIG and then SETUP).
-Use
or
range keys to scroll through the menu items.
-Press Enter to select or Exit to quit without changing power-on setup.
If DDC language is selected, user setups cannot be saved as power-on setup.
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......