Range, Units, Digits, Rate, and Filters
4-3
Autorange limits
Search time for amps can be reduced by setting upper and/or lower autorange limits. For
example, if you know the maximum input will be around 1µA, set the upper current range
limit to 2µA. This eliminates the 20µA, 200µA, 2mA, and 20mA ranges from the search,
thereby increasing the range change speed. Should the input exceed 2.1µA, the OVR-
FLOW message will be displayed.
Perform the following steps to set upper and/or lower autorange limits.
1.
Press CONFIG key (CONFIGURE: will be displayed).
2.
Display the desired limit (UPPER or LOWER):
a.
Press the RANGE
key to display the present UPPER range limit.
b.
Press the RANGE
key to display the present LOWER range limit.
3.
Scroll through the available range limits using the
and
range keys.
4.
Press ENTER when the desired range is flashing.
NOTE
If you attempt to select an incompatible range limit, it will be ignored and TOO
LARGE or TOO SMALL will be displayed briefly. For example, if the lower
range limit is 20µA, trying to set the upper limit to 2µA will result in the TOO
SMALL error.
Units
Readings can be displayed using engineering (ENG) units (i.e. 1.236 m
Α
) or scientific
(SCI) notation (i.e. 1.236E-03
Α
). Perform the following steps to change the units setting:
1.
Press MENU key.
2.
Scroll down to the UNITS item using the
or
range keys (UNITS: will be flash-
ing).
3.
Press ENTER to select setting (ENG or SCI will be flashing).
4.
Use the
or
key to display the desired units setting.
5.
Press ENTER.
NOTE
1. The units setting can only be changed from the front panel (no remote
operation).
2. Scientific notation provides more resolution on small values than
engineering units.
Digits
The DIGITS key sets display resolution for Model 6485. Display resolution can be set
from 3 to 6 digits. This single global setting affects display resolution for all measure-
ment ranges.
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......