Model 6485 Picoammeter Instruction Manual
Range, Units, Digits, Rate, and Filters
4-11
Digital filter control
The AVG key is a toggle key: it will either enable the digital filter (displays AVERAGE
ON) or disable the digital filter (displays AVERAGE OFF). To configure the median filter:
NOTE
For the following procedure, use the
and
keys and the RANGE
and
keys to set values. The
and
keys provide cursor control, while the
and
keys increment and decrement the value.
To return to the previously set value and cancel configuration, press EXIT
instead of ENTER. Values change immediately when the ENTER key is pressed.
1.
Press the CONFIG key.
2.
Press the AVG key. The present number of reading conversions to average (filter
count) will be displayed (flashing).
3.
Set filter count (1 to 100):
•
Use the range keys
or
to display the desired filter count value at the
RDGS prompt.
•
Press ENTER to set.
4.
Set filter type (REPEAT or MOVING AV):
•
Use the range keys
or
to display the desired filter type at the TYPE:
prompt.
•
Press ENTER to set.
5.
Enable or disable advanced filter (YES or NO).
•
Use the range keys
or
to display the desired state at the ADVANCED?:
prompt.
•
Press ENTER to set.
6.
If the advanced filter is enabled (ADVANCED?:YES), enter the noise tolerance
level at the NTOL: prompt (0–105.0000%).
•
Use the range keys
,
,
, and
, to display the desired value.
•
Press ENTER to set.
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......