5-6
Relative, mX+b, m/X+b (reciprocal), and log
Model 6485 Picoammeter Instruction Manual
2.
Key in a scale factor value. The
and
keys control cursor position and the
and
range keys increment and decrement the digit value. To change range, place the
cursor on the range symbol and use the
and
keys. With the cursor on the polar-
ity sign, the
and
keys toggle polarity.
NOTE
Range symbols are defined in
3.
Press ENTER to enter the M value and display the offset (B) value:
B: +00.000000 P (factory default)
4.
Key in the offset value.
5.
Press ENTER to set the B value and display the one-character UNITS designator:
UNITS: X (factory default)
NOTE
The configuration for mX+b calculations consists of a units designator, a value
for M, and a value for B. This configuration is used for both the mX+b and the
m/X +b calculations. Therefore, changing either configuration (of the mX+b or
the m/X+b calculation) also changes the other calculation’s configuration.
6.
To change the units designator (default is “X”, press the
cursor key, and use the
and
range keys. The character can be any letter in the alphabet (A through Z).
7.
Press ENTER. The MATH annunciator and the units designator will turn on, and
the result of the calculation will be displayed. Note that the calculation will be
applied to all measurement functions.
Logarithmic
This calculation converts input readings to logarithm base 10 values. The calculation is
performed as follows:
where: X is the input reading
y is the logarithmic result
For example: Assume that exactly 1mA is being measured by the Model 6485.
NOTE
This calculation uses the absolute value of the normal input reading as the log
of a negative number cannot be computed.
10
X
log
Y
=
1.000000mA
10
log
3
–
=
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......