6-6
Buffer
Model 6485 Picoammeter Instruction Manual
A) TRACe:FREE?
Two values, separated by commas, are returned. The first value indicates how many bytes
of memory are available, and the second value indicates how many bytes are reserved to
store readings.
B) TRACe:FEED <name>
Name parameters:
•
SENSe — Raw input readings are stored in the buffer.
•
CALCulate1 — The results of the mX+b, m/X+b, or log calculation are stored in
the buffer. See
for information on mX+b, m/X+b, or log.
•
CALCulate2 — Test limit or Rel readings are stored in the buffer. See
information on limit tests.
C) TRACe:FEED:CONTrol <name>
Name parameters:
•
NEXT — Enables the buffer and turns on the asterisk (*) annunciator. After the
buffer stores the specified number of readings, the asterisk annunciator turns off.
•
NEVer — Disables the buffer.
D) TRACe:TSTamp:FORMat <name>
Name parameters:
•
ABSolute — Each timestamp is referenced to the first reading stored in the buffer.
•
DELTa — Timestamps provide the time between each buffer reading.
•
The timestamp data element can be included with each buffer reading (see Ref F).
E) TRACe:DATA?
1.
The response message will include one to four data elements for each stored read-
ing. Use the FORMat:ELEMents command (see Ref F) to specify the elements.
2.
Reading an empty buffer will result in the “ERROR: -230” display message.
3.
Buffer data can be sent in the binary format. (See
command summary,” on page 5 for details).
F) FORMat:ELEMents <list>
List parameters:
•
READing — Includes the buffer reading in each data string.
•
UNITs — identifies amps or math units.
•
TIME — Includes the timestamp for each reading. Timestamp can be in the abso-
lute or delta format (see Ref D).
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......