Model 6485 Picoammeter Instruction Manual
Remote Operation
9-7
Primary address
Model 6485 ships from the factory with a GPIB address of 14. When the instrument pow-
ers up, it momentarily displays the primary address. You can set the address to a value of
0-30. Do not assign the same address to another device or to a controller that is on the
same GPIB bus.
Usually controller addresses are 0 or 21, but see the controllers instruction manual for
details. Make sure the address of the controller is the same as that specified in the control-
lers programming language. To make sure the units interface is properly selected and con-
figured, or to check or change the GPIB address, refer to
.
General IEEE-488 bus commands
Commands and associated statements
General commands are those commands, such as DCL, that have the same general mean-
ing regardless of the instrument.
lists the general bus commands.
REN (remote enable)
The remote enable command is sent to Model 6485 by the controller to set up the instru-
ment for remote operation. Generally, the instrument should be placed in the remote mode
before you attempt to program it over the bus. Simply setting REN true does not actually
place the instrument in the remote state. You must address the instrument to listen after
setting REN true before it goes into remote.
Note that the instrument does not have to be in remote to be a talker.
Table 9-1
General bus commands
Command
Effect on Model 6485
REN
Goes into remote when next addressed to listen.
IFC
Reset interface; all devices go into talker and listener idle states.
LLO
LOCAL key locked out.
GTL
Cancel remote; restore front panel operation for Model 6485.
DCL
Returns all devices to known conditions.
SDC
Returns Model 6485 to known conditions.
GET
Initiates a trigger.
SPE, SPD
Serial polls Model 6485.
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......