Model 6485 Instruction Manual
Performance Verification
15-5
Verification limits
The verification limits stated in this section have been calculated using only Model 6485
one-year accuracy specifications, and
they do
not
include test equipment uncertainty
. If a
particular measurement falls outside the allowable range, recalculate new limits based on
both Model 6485 specifications and corresponding test equipment specifications.
Example reading limits calculation
As an example of how verification limits are calculated, assume you are testing the 20mA
range using a 20mA input value. Using the Model 6485 20mA range accuracy specifica-
tion of ±(0.1% of r 1
µ
A), the calculated reading limits are:
Reading limits = 20mA ± [(20mA
×
0.1%) + 1
µ
A]
= 20mA ± (0.02mA + 0.001mA)
= 20mA ± 0.021mA
= 19.979mA to 20.021mA
Calibrator voltage calculations
When verifying the 2nA-2
µ
A current ranges, you must calculate the actual calibrator volt-
ages from the desired current values and the characterized Model 5156 Calibration Stan-
dard resistor values.
Calibrator voltages required for verification currents are calculated as follows:
V = IR
Where:
V = required calibrator voltage
I = verification current
R = actual standard resistor value
For example, assume you are testing the 20nA range using an actual 100.5M
Ω
standard
resistor value. The actual calibrator voltage is: 20nA
×
100.5M
Ω
= 2.01V.
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......