Model 6485 Instruction Manual
Calibration
16-3
Calibration considerations
When performing the calibration procedures:
•
Make sure that the test equipment is properly warmed up and connected to the
Model 6485 INPUT jack.
•
Always allow the source signal to settle before calibrating each point.
•
Do not connect test equipment to the Model 6485 through a scanner or other
switching equipment.
•
If an error occurs during calibration the Model 6485 will generate an appropriate
error message.
WARNING
The maximum safe voltage between picoammeter LO and chassis
ground (common mode voltage) is 42V. The Model 6485 does not inter-
nally limit the LO to chassis voltage. Exceeding 42V can create a shock
hazard. The LO to chassis breakdown voltage is 500V. Exceeding this
voltage may cause damage to the instrument.
CAUTION
Maximum continuous input voltage is 220V DC. Exceeding this value
may cause instrument damage.
Calibration cycle
Perform verification at least once a year to ensure the unit meets or exceeds its specifica-
tions. Calibrate if necessary.
Recommended calibration equipment
lists the recommended equipment for the calibration procedures. You can use
alternate equipment, but keep in mind that test equipment uncertainty will affect calibra-
tion accuracy. Calibration equipment should have accuracy specifications at least four
times better than corresponding Model 6485 specifications. Note, however, that the rec-
ommended calibrator listed in
does not meet this requirement for 20
µ
A output.
Summary of Contents for 6485
Page 20: ......
Page 24: ......
Page 44: ...1 20 Getting Started Model 6485 Picoammeter Instruction Manual ...
Page 66: ...3 6 Measurements Model 6485 Picoammeter Instruction Manual ...
Page 138: ...9 20 Remote Operation Model 6485 Picoammeter Instruction Manual ...
Page 159: ...11 CommonCommands ...
Page 164: ...11 6 Common Commands Model 6485 Picoammeter Instruction Manual ...
Page 165: ...12 SCPISignalOriented MeasurementCommands ...
Page 190: ...14 12 SCPI Reference Tables Model 6485 Picoammeter Instruction Manual ...
Page 214: ...16 14 Calibration Model 6485 Instruction Manual ...
Page 219: ...A Specifications ...
Page 221: ...B StatusandErrorMessages ...
Page 227: ...C GeneralMeasurement Considerations ...
Page 233: ...D DDCEmulationCommands ...
Page 247: ...E ExamplePrograms ...
Page 250: ...E 4 Example Programs Model 6485 Picoammeter Instruction Manual ...
Page 251: ...F IEEE 488BusOverview ...
Page 266: ...F 16 IEEE 488 Bus Overview Model 6485 Picoammeter Instruction Manual ...
Page 267: ...G IEEE 488andSCPI ConformanceInformation ...
Page 311: ......